Microcatheters

ASAHI Tellus

Indications for Use

This microcatheter is a medical device intended for angiography and/or infusion of various substances including diagnosis, embolization and treatment in the peripheral vasculature using an appropriate guide catheter and guide wire. The microcatheter is not intended for use in neurovasculature, coronary arteries and carotid arteries.

Key Specifications

  • Catheter shaft reinforced with braid wires to enhance pushability and maintain patency of the inner lumen
  • Radiopaque marker fixed on the distal end for location during angiography
  • Distal portion flexible with straight and angled shapes for improved trackability in tortuous vessels
  • Hydrophilic coating on outer surface for smooth transition
  • Outer Diameter (OD): 0.41 mm (0.016")
  • Max Guide Wire Diameter: Distal 0.58 mm (1.7 Fr, 0.023"), Proximal 0.94 mm (2.8 Fr, 0.037")
  • Minimum Inner Diameter (ID): 1.05 mm (0.041")
  • Effective Length: 105 cm, 125 cm, 150 cm
  • Tip Shape: Straight, Angled
  • Sterile single-use device
  • Sterilization: Ethylene Oxide
  • Accessories: Stylet, syringe, and RHV (rotating hemostasis valve), or stylet, syringe, inserter, and hemostasis valve
  • Guide Catheter Compatibility: Minimum ID 1.05 mm (0.041")
  • Tested for: Appearance/Dimensions/Tip Shape, Corrosion Resistance, Peak Tensile Strength, Tip Flexibility, Liquid Leakage, Radio-detectability, Air Leakage, Burst Pressure, Flow Rate, Power Injection, Kink Resistance, Slidability, Connector, Coat Integrity/Particulate Evaluation, Torque Strength (all passed)

Description

The ASAHI Tellus Microcatheter is a sterile single-use device designed for use in the peripheral vasculature. It consists of a catheter shaft reinforced with braid wires to enhance pushability and maintain patency of the inner lumen. A radiopaque marker is fixed on the distal end to facilitate location during angiography. The distal portion is flexible and available in straight and angled shapes for improved trackability in tortuous vessels. A hydrophilic coating is applied on the outer surface for a smooth transition.