N
NeuroDirectory
Aspiration Catheters

072 Aspiration System

Indications for Use

072 Aspiration Catheter: As part of the 072 Aspiration System, the 072 Aspiration Catheter with a compatible suction pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment. Aspiration Tubing: As part of the 072 Aspiration System, the Aspiration Tubing is intended to connect the 072 Aspiration Catheter to a compatible suction pump.

Key Specifications

  • Single lumen, variable stiffness catheter
  • Hydrophilic coating to reduce friction during use
  • Radiopaque markers on distal end for angiographic visualization
  • Luer hub on proximal end for flushing and aspiration
  • Length: 132 cm
  • Inner Diameter (ID): 0.072 in
  • Distal Outer Diameter (OD): 0.0855 in
  • Proximal OD: 0.0855 in
  • Coating Length: 28 cm +/- 4 cm
  • Tip Configuration: Straight
  • Hub Material: Polycarbonate
  • Liner Material: PTFE
  • Catheter Shaft Extrusions: Nylon, Pebax
  • Marker Band Material: Platinum (90%) / Iridium (10%), Tantalum
  • Reinforced Shaft: Stainless Steel (SS) Coil
  • Strain Relief Material: Polyolefin
  • Delivery Tool Materials: Pebax, Tecoflex 93A, Tungsten 75% Filler
  • Delivery Tool Dimensions: ID 0.026 in, OD 0.042 in, Extension Length 10 cm
  • Aspiration Tubing ID: 0.110 in
  • Aspiration Tubing Length: 95 in
  • Flow Control Mechanism: Flow Control Switch
  • Shelf Life: 6 months
  • Sterilization Method: Ethylene Oxide
  • Sterility Assurance Level (SAL): 10^-6
  • Includes: 072 Aspiration Catheter, Delivery Tool, Rotating Hemostasis Valve, Flow Switch
  • Aspiration Tubing provided separately
  • Used with compatible suction pump

Description

The 072 Aspiration Catheter is a single lumen, variable stiffness catheter with a hydrophilic coating to reduce friction during use. It includes radiopaque markers on the distal end for angiographic visualization and a Luer hub on the proximal end for attachments for flushing and aspiration. The Delivery Tool is an optional accessory and should be removed prior to aspiration. The package includes the 072 Aspiration Catheter, Delivery Tool, Rotating Hemostasis Valve, and Flow Switch. The Aspiration Tubing is provided in a separate package. The catheter is used in conjunction with a compatible suction pump connected via the Aspiration Tubing.