MIVI Q Distal Access Catheter

Indications for Use

The MIVI Q Distal Access Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.

Key Specifications

  • Single-lumen, variable stiffness catheter
  • Radiopaque markers on distal and proximal ends for angiographic visualization
  • Hydrophilic coating on catheter shaft to reduce friction
  • Proximal portion consists of stainless-steel control (push) wire
  • Introduced via 8F guide catheter or 6F guide sheath over guidewire or microcatheter
  • Compatible with Q Docking Station for insertion and extraction through hemostasis valve
  • Pin vise used to advance the catheter
  • Braided construction for 5F and 6F sizes; coiled for 3F and 4F sizes
  • Straight tip
  • Two radiopaque markers (distal and proximal)
  • Available sizes: 3F, 4F, 5F, 6F
  • Total lengths: 3F - 154 cm, 4F - 141 cm, 5F - 136 cm, 6F - 136 cm
  • Extensible lengths with guide (90-95 cm): 3F - 148 cm, 4F - 135 cm, 5F - 130 cm, 6F - 130 cm
  • Control wire lengths: 3F - 108 cm, 4F/5F/6F - 104 cm
  • Compatible guidewire diameters: 0.018" for 3F and 4F, 0.035" for 5F and 6F
  • Accessories include 9F Rotating Hemostasis Valve Y-Connector and Q Docking Station
  • Sterilization method: Ethylene Oxide (EO)
  • Shelf life: 3 years
  • Patient contacting materials identical to predicate device

Description

The MIVI Q Distal Access Catheter (Q Catheter) is a single-lumen, variable stiffness catheter with radiopaque markers on the distal and proximal end of the catheter portion for angiographic visualization. The catheter shaft has a hydrophilic coating to reduce friction during use. The proximal portion of the catheter is a stainless-steel control (push) wire. The Q Catheter may be introduced via an 8F guide catheter/6F guide sheath and over a guidewire/microcatheter into the arterial vasculature until the desired vessel is reached. The Q Docking Station may be used to facilitate insertion and extraction of the Q Catheter through a hemostasis valve attached to the 8F guide catheter/6F guide sheath. The pin vise may be used to advance the catheter.