Aspiration Catheters

Next Generation Aspiration Catheter; Balt Aspiration Tubing Set

Indications for Use

{"Next Generation Aspiration Catheter":"The Next Generation Aspiration Catheter with a compatible aspiration pump and Balt Aspiration Tubing Set is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV-tPA) or who fail IV-tPA therapy are candidates for treatment.","Balt Aspiration Tubing Set":"The Balt Aspiration Tubing Set is intended to connect the Next Generation Aspiration Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow."}

Key Specifications

  • Single-lumen, coil reinforced, variable stiffness composite catheter
  • Hollow cylindrical tube with standard luer fitting
  • Wall constructed using metal coils and medical grade polymers
  • Hydrophilic coating on distal portion for lubricious surface
  • Radiopaque marker at distal tip for fluoroscopic visualization
  • Offered in various working lengths: 120 cm, 128 cm, 132 cm, 137 cm
  • Nominal inner diameters: 0.071 inch, 0.074 inch
  • Nominal outer diameters: proximal 0.086 inch, distal 0.085 inch
  • Introducer sheath provided for insertion support
  • Sterile, non-pyrogenic, single use only
  • Balt Aspiration Tubing Set: length 114 inch, inner diameter 0.110 inch ± 0.003 inch
  • Tubing made of medical grade polymers with luer fitting and slip fit connector
  • Tubing includes flow control valve
  • Sterilization: Ethylene Oxide (EO), SAL 10-6
  • Shelf life: 12 months for catheter, 6 months for tubing set
  • Intended for use with guidewires, guide sheaths, and compatible aspiration pump

Description

The Next Generation Aspiration Catheter is a single-lumen, coil reinforced, variable stiffness composite catheter that facilitates removal of thrombus from the neurovasculature when connected to a compatible aspiration pump and the Balt Aspiration Tubing Set. It is comprised of a hollow cylindrical tube with a standard luer fitting, constructed using metal coils and medical grade polymers, featuring a hydrophilic coating on the distal portion and a radiopaque marker at the distal tip. Offered in various lengths and diameters, it is provided sterile for single use with an introducer sheath. The Balt Aspiration Tubing Set is a hollow cylindrical tube bonded to a luer fitting and slip fit connector, made of medical grade polymers, also sterile and single use.