Indications for Use

Millipede 070 Aspiration Catheter: The Millipede 070 Aspiration Catheter, with the Perfuze Aspiration Tube Set and a compatible aspiration pump, is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. Perfuze Aspiration Tube Set: The Perfuze Aspiration Tube Set is indicated to connect the Millipede 070 Aspiration Catheter to a compatible aspiration pump.

Key Specifications

  • Single lumen, reinforced, variable stiffness catheter
  • Distal segment with hydrophilic coating for navigation
  • Radiopaque marker at distal end for fluoroscopic visualization
  • Rotating hemostasis valve (RHV) on hub for hemostasis during infusion and device insertion
  • Tapered, soft, flexible distal tip
  • Braid and coil reinforcement with ribbed internal surface at distal end
  • Inner diameter: Distal 0.070 in, Proximal 0.069 in
  • Outer diameter: Distal 0.0835 in, Proximal 0.0865 in
  • Effective length: 136 cm
  • Sterile, single-use device
  • Used with Perfuze Aspiration Tube Set: 100 in length, Tubing ID 0.110 in, rotating male luer-lock connector, suction connector, pinch clamp for vacuum control
  • Materials: Polymers and metals commonly used in medical devices
  • Sterilization: Ethylene Oxide (EO) with SAL 1x10-6
  • Shelf life: 12 months for catheter, 24 months for tube set

Description

The Millipede 070 Aspiration Catheter is a sterile single-use device consisting of the catheter and a rotating hemostasis valve (RHV). The catheter is a single lumen, reinforced, variable stiffness catheter. The distal segment has a hydrophilic coating for navigation through the vasculature. The catheter has a radiopaque marker located at its distal end for visualization under fluoroscopy. The RHV is assembled onto the hub of the catheter and is used to maintain hemostasis during infusion of saline and contrast agent and insertion of devices into the Millipede 070 Aspiration Catheter. The Perfuze Aspiration Tube Set is a sterile, single-use device consisting of a single flexible braided tube with a rotating male luer-lock connector on the distal end, a suction connector on the proximal end, and a pinch clamp. For the aspiration source, the Millipede 070 Aspiration Catheter is used in conjunction with a compatible aspiration pump connected using the Perfuze Aspiration Tube Set along with a legally marketed canister and accessories kit.