Balloon Guide Catheters

Modified FlowGate Balloon Guide Catheter

Indications for Use

The Concentric Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

Key Specifications

  • Coaxial-lumen, braid-reinforced, variable stiffness catheter
  • Designed for peripheral and neurovascular systems
  • Radiopaque marker on distal end for angiographic visualization
  • Compliant balloon mounted on distal end for temporary vascular occlusion
  • Bifurcated luer hub on proximal end for flushing, inflation, and aspiration
  • Outer/Inner Jacket: Pebax with revisions to transition lengths and material durometers
  • Distal Tip: Pebax
  • Balloon Material: Silicone
  • Braid: Stainless Steel
  • Braid distal end securement: Cyanoacrylate with different viscosity
  • Marker Band: Platinum/Iridium
  • Catheter Hub: Polyurethane
  • Strain Relief: Polyolefin with additional Pebax strain relief
  • Labeled Shaft Inner Diameter (nominal): 0.084 in
  • Labeled Shaft Outer Diameter (nominal): 0.106 in
  • Maximum outer diameter along effective length: 0.1080 in
  • Effective Lengths: 85 cm and 95 cm
  • Distal Tip length: 0.75 mm
  • Radiopaque Marker Length (nominal): 0.020 in
  • Accessories: Dilator, Rotating Hemostasis Valve, Tuohy Borst Valve with sideport, Peel Away Sheaths, Luer-Activated Valve, Extension tubing
  • Packaging: Polyethylene Tube and HDPE Packaging Card, Tyvek/LLDPE Pouch
  • Sterilization Method: Ethylene Oxide
  • How Supplied: Sterile, Single Use
  • Biocompatible: Non-sensitizer, Non-cytotoxic, Non-irritant, No acute systemic toxicity, Non-pyrogenic, Non-hemolytic, Minimal complement activation potential, No significant thrombosis

Description

The Modified FlowGate Balloon Guide Catheter is a coaxial-lumen, braid-reinforced, variable stiffness catheter designed for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. A radiopaque marker is included on the distal end for angiographic visualization. A compliant balloon is mounted on the distal end to provide temporary vascular occlusion during angiographic procedures. A bifurcated luer hub on the proximal end allows attachments for flushing, inflation and aspiration. Modified FlowGate Balloon Guide Catheter dimensions and maximum recommended balloon inflation volume are indicated on product label.